We present a portable four-channel atomic magnetometer array operating in the spin-exchange relaxation-free regime. The magnetometer array has several design features intended to maximize its suitability for biomagnetic measurement, specifically foetal magnetocardiography, such as a compact modular design and fibre-coupled lasers. The modular design allows the independent positioning and orientation of each magnetometer. Using this array in a magnetically shielded room, we acquire adult magnetocadiograms. These measurements were taken with a 6-11 fT Hz(-1/2) single-channel baseline sensitivity that is consistent with the independently measured noise level of the magnetically shielded room.
PMID: 22504066