Non-contact determination of parasympathetic activation induced by a full stomach using microwave radar.

Author: Gotoh S, Suzuki S, Imuta H, Kagawa M, Badarch Z, Matsui T.
Affiliation:
Department of Management Systems Engineering, Faculty of System Design, Tokyo Metropolitan University, 6-6 Asahigaoka, Hino, Tokyo, 191-0065, Japan.
Conference/Journal: Med Biol Eng Comput.
Date published: 2009 Jul 5
Other: Word Count: 145


In order to evaluate parasympathetic activation which causes driving errors, without placing any burden on the monitored individuals, we conducted a non-contact parasympathetic activation monitoring through the back of a chair using a compact 24-GHz microwave-radar. We measured the high-frequency (HF, 0.15-0.4 Hz) power spectrum of heart rate variability (HRV) which reflects parasympathetic activation, induced by a full stomach. All participants had a large all-you-can-eat meal with beverages for lunch within 20 min. Before and after the large meals for durations of 10 min, the non-contact measurement was conducted for seven healthy male volunteers (mean age: 23 +/- 1-year-old). In both non-contact (microwave radar) and contact (ECG as a reference) measurement, HF shows similar variations before and after large meal. Large meal significantly (p < 0.05) increased non-contact-derived HF from 1,026 +/- 510 to 1,893 +/- 613 ms(2) (922 +/- 628 to 1,861 +/- 940 ms(2), p < 0.05). This technique allows parasympathetic activation monitoring for safety precautions.