PURPOSE:
This paper intends to demonstrate the feasibility of truly combined PET/CT imaging and addresses some of the major challenges raised by this dual modality approach. A method is proposed to retrieve maximum accuracy out of limited resolution computed tomography (CT) scans acquired with positron emission tomography (PET) detectors.
METHODS:
A PET/CT simulator was built using the LabPET™ detectors and front-end electronics. Acquisitions of energy-binned data sets were made using this low spatial resolution CT system in photon counting mode. To overcome the limitations of the filtered back-projection technique, an iterative reconstruction library was developed and tested for the counting mode CT. Construction of the system matrix is based on a preregistered raster scan from which the experimental detector response is obtained. PET data were obtained sequentially with CT in a conventional manner.
RESULTS:
A meticulous description of the system geometry and misalignment corrections is imperative and was incorporated into the matrix definition to achieve good image quality. Using this method, no sinogram precorrection or interpolation is necessary and measured projections can be used as raw input data for the iterative reconstruction algorithm. Genuine dual modality PET/CT images of phantoms and animals were obtained for the first time using the same detection platform.
CONCLUSIONS:
CT and fused PET/CT images show that LabPET™ detectors can be successfully used as individual X-ray photon counting devices for low-dose CT imaging of the anatomy in a molecular PET imaging context.